enPulsPro

Progressive therapy system with comfortable housing for all needed accessoriesenPulsPro – a Radial Shockwave Therapy System with ballistic, high energy pulses. For treatment of biological tissues and structures in various pathologies and medical conditions. An unique, easy-to-use system offering shockwave therapy access to all therapists.

Description

Modern technology, innovatively designed
The new enPuls user interface is highly intuitive and easy-to-use. All parameters are optimally designed for a clear layout. Ready for your therapy.
Zimmer MedizinSysteme has created something really special with this new generation of enPuls products. An entire new user experience. New capacitive touch screen technology provides even simpler operation and application.
Higher treatment comfort and time saving by the opportunity to connect two handpieces with different applicators at the same time.
Radial Shockwaves – Unique and powerful technologyhandstueck_schulter
Radial shockwave therapy (RSWT) is a method for the treatment of superficial orthopedic disorders developed over the past 20 years. A projectile accelerated in the handpiece generates a mechanical pressure wave, which is transferred to the human body by an applicator head and which radiates out in the tissue.The mechanical energy is absorbed by the tissue and decreases as the distance from the application site increases. The mechanical stress triggers reactions in the tissue that have a positive influence on a wide range of orthopedic and neurological conditions.The method is used by therapists all over the world with great success. With the enPulsPro, Zimmer MedizinSysteme now offers a system that bundles all the advantages of the application in a compact form with the lowest downstream and operating costs that is easy to operate and apply with maximum flexibility and mobility.
Appliance
enPulsPro is ideal as a system for radial shockwave therapy, for superficial orthopedic problems such as:
  • Insertion tendinopathy
  • Tendon problems
  • Chronic inflammation
  • Hypertonic muscle systems
  • Myofascial syndromes
  • Hemotoma treatments
softshot – pulse

enPulsPro generates a special form of shockwave with a relatively heavy projectile and a comparatively low impact speed—the enPulsPro softshot shockwave.
The energy is primarily generated by the volume of the pulse and not by the high amplitude and the extremely short rise characteristic of comparable compressed-air devices.
This requires that the shockwaves with rise times of approx. 3.5 µs and the typical energy settings also generated by compressed-air devices pass through the tissue without any change in the form of the wave, because the stress is within the moment of inertia and the module of elasticity of the tissue.
Different from focused shockwaves, which have rise times less than 100 ns and which results in a desired gradation of the wave in the tissue causing cavitation and tissue destruction, with radial shockwaves it is not the exponential effect but simply the amount of energy that triggers the reactive processes.
The slower rise of the wave (approx. 13.5 µs) and the significantly lower peak amplitude mean that enPulsPro does not generate the typical whiplash effect of the shockwave, make it easier for the patient to tolerate.
Energy applied—pressure comparison

Comprehensive comparative tests in the laboratory have demonstrated that enPulsPro with equivalent energy settings releases kinetic energy via the applicator comparable to that of a compressed-air-driven system, in which the air pressure applied to accelerate the projectile is the dosage.

The similarity is shown in the following table

enPulsPro         Compressed-air RSWT-System
60 mJ 1 bar
90 mJ 2 bar
120 mJ 3 bar
185 mJ 5 bar

Because the kinetic energy is applied to the tissue with similar-sized applicators, the standard units such as mJ/mm² (energy density) can be used because of the relation.

Dosage information from treatment protocols can thus be used 1:1.

Low maintenance and long life

The special technology of the enPulsPro and the softshot shock source result in low maintenance costs of the system as well as the longest handpiece and shockwave generator service life compared to compressed-air systems.

Zimmer MedizinSysteme offers a minimum guarantee of 2.000.000 shocks per generator, corresponding to about 1.000 treatments. The generator should be replaced if there is a significant reduction in performance (shock power is reduced, misfire, etc.). This is generally noticed long after the minimum guarantee period has expired.

VAS-Scale

The assessment of pain intensity with the integrated VAS scale provides a precise and traceable view on the progress of the therapy to both therapist and patient

Technical data

Technology
Compressor free ballistic radial shockwave therapy-system with electromagnetic generator as projectile accelerator

Power Levels / Energy

  • 60 to 185 mJ (equivalent to 1-5 bar)
  • Can be precisely set in 10 mJ intervals (depending on the frequency of 60 mJ to max. 185 mJ)

Modes

  • Frequencies from 1 to 22 Hz
  • 3 Burst modes (4 / 8 / 12 pulses)

Programmes
7 preset programmes, adjustable

Protocols
More than 25 illustrated preset treatment recommendations

Controls

  • 12” Colour touch-screen for all software operations
  • Rotary control for energy level and for frequency
  • Handpiece operation with multi-directional footswitch

Memory / Update

  • SD-card for indication menu memory
  • favourites and adapted programmes (120 places)
  • Error log / Firmware-Update

Dimensions
20 x 35 x 30 cm (L/W/H)

Weight
3.8 kg (without handpiece)

Shockwave Applicator Hand Piece

  • Ergonomic, with anodized aluminium casing and fan cooling
  • Dimensions 23 cm length, 5 cm diameter (max.)
  • Weight approx 0.85 kg (without cable)
  • Lifetime Warranty of minimum 2.000.000 shocks
  • Maintenance Only necessary when performance drops, not obligatory at a
  • certain amount of shocks. Exchange of generator by service
  • Applicator heads Diameters of 6 / 15 (2 pieces) / 25 mm
  • Tool-free exchange possibility
  • Lifetime 150.000 shocks per applicator head

Power Consumption
100 – 240 V / 50/60 Hz, max. 250 VA
220 V / 60 Hz

Optional

  • Trolley SysCart incl. holder for 2 handpieces and lotion
  • Tray for applicators
  • 2nd Handpiece
  • Swivel base

Intended use and indications may not be valid for the USA.